Grade B microbial action limits fit on one line: 10 CFU/m³ of air, 5 per settle plate over 4 hours, 5 per contact plate, 5 per glove print.
Those are the maximum action limits in Table 6 of EU GMP Annex 1 (2022). A Grade B room whose settle plates read 3 or 4 every week is inside them and is still telling you something.
Annex 1 expects alert levels below the action limits, set from your own data. Where a room's history is mostly zero recoveries, a repeated count of 1 or 2 at the same location is worth an alert long before anything reaches 5. That only works if someone reads the results as a trend: the same organism, the same location, the same shift.
A new facility has no history to set them from. Its early alert levels are a judgement, written down with the reason and revised once enough data exists, and the PQ and early routine monitoring should be planned as the baseline for that revision.
Settle plates carry a time limit too. The 5 CFU applies per plate exposed for up to 4 hours, so a longer session needs plate changes, and a gap between plates is a gap in the record.
HVAC cannot fix aseptic behaviour. When counts cluster near a door, a return grille or one side of a line, though, the trend is usually pointing at airflow, and that belongs in the Contamination Control Strategy.