Five daily checks on a negative-pressure isolation room

  • Life Sciences
  • 1 min read

At the gap under an isolation room door, a smoke tube shows which way the air moves, whatever the pressure monitor says.

A monitor that has read −2.5 Pa for a month may be right. It may also have stopped seeing the room: a kinked sensing tube, a blocked port, a transmitter that repeats its last value.

The CDC guidelines on environmental infection control ask for continuous negative pressure of 2.5 Pa against the corridor, monitored periodically, preferably daily, with audible manometers or smoke tubes at the door or a permanently installed visual monitor, and documented. A daily round, with the door closed, takes a few minutes:

  1. The monitor reads inside its limits and the alarm has not been silenced. Open the door for a moment and a live monitor moves.
  2. Smoke from a tube, or a strip of tissue, held at the gap under the closed door is drawn into the room.
  3. The door closes by itself and the latch catches when the door is let go.
  4. Both anteroom doors are shut.
  5. The exhaust fan shows running at the panel, and the result goes in the log with a time and initials.

Step 2 tends to be the first one dropped once a digital monitor is fitted.

Who signs the daily check on your isolation rooms: nursing, or engineering?

Industries

Related

More on this subject

Five things to read on a calibration certificate

Most calibration certificates are filed unread. Check that the laboratory is accredited for this measurement, that tag, serial number and calibrated points match the instrument and its range, read as-found before as-left, compare the uncertainty with the tolerance, and look for traceability and the decision rule behind any pass statement.

InsightRead

Fixed-speed FFUs slow down as their HEPA loads

A fixed-speed fan filter unit loses airflow as its HEPA filter loads. Set at 0.45 m/s, it has 20% to lose before the 0.36 m/s bottom of the Annex 1 guidance range. Constant-airflow EC units hold velocity if they are selected at the final pressure drop, and their speed trend doubles as a filter-loading record.

InsightRead

Four checks at the basic-design gate

Before detail design opens, a GMP project should have frozen room data sheets and pressure cascade, loads that match the equipment list, a URS for each critical system and owned open items. A cascade change after the ducts are drawn reaches the duct sizes, terminal filters, AHU selection and every neighbouring pressure step.

InsightRead