A double-door autoclave sterilises what is inside the load. A VHP pass-through only decontaminates the surfaces the vapour can reach.
That difference decides which route each item takes into the Grade A zone of a fill-finish line. There are usually three, plus the depyrogenation tunnel that brings in the glass.
- Double-door autoclave, built through the wall from the preparation side into the Grade B side. For items that tolerate moist heat: change parts, tools, stopper bowls, wrapped garments. Steam penetrates the wrapping, so the load arrives sterile inside and out. Its doors are interlocked and the cycle is validated with load patterns and biological indicators.
- VHP pass-through, for items that cannot take steam, such as electronics and sealed packs. It treats surfaces only. Occluded surfaces, stacked items and absorbent materials are its weak points, so biological indicators go where the vapour struggles to reach.
- Rapid transfer port on an isolator or RABS. Pre-sterilised items, typically in gamma-irradiated bags or containers, dock and open straight into Grade A without passing through the room. The weak spot is the seal ring exposed when the doors open, so the docking sequence is written into the procedure.
Each route belongs in the CCS with its validation evidence.