The accelerated stability condition: 6 months at 40 °C / 75% RH

  • News
  • 1 min read

Six months at 40 °C and 75% RH is usually the first evidence a dossier has about shelf life.

The condition comes from the ICH stability guideline, Q1A(R2). Its general case sets the accelerated study at 40 °C ± 2 °C and 75% RH ± 5% RH, with a minimum of 6 months of data at submission.

If long-term studies run at 25 °C/60% RH and a "significant change" appears at any time in those six months, additional testing at the intermediate condition should be conducted. A 5% change in assay from its initial value is one of the listed changes.

For the people who run the chamber, the tolerances are the specification. The guideline's glossary says the equipment should be capable of controlling within those ranges, and that actual temperature and humidity should be monitored. Excursions beyond the tolerances for more than 24 hours should be described in the study report and their effect assessed.

A weekend is longer than 24 hours. A humidifier that fails on Friday evening and is found on Monday morning is in the stability report.

How long can an alarm on your stability chamber wait before someone answers it?

Related

More on this subject

The engineering section of the Site Master File

The Site Master File is read when an inspection is planned, and section 4 of the content list in the EU Explanatory Notes is engineering content: layouts with room classification and pressure differentials, HVAC principles, water schematics, other utilities and the major equipment list. Each appendix can carry its own effective date, so check them after every project.

InsightRead

The first 90 days of a GMP maintenance takeover

Taking over HVAC maintenance in a running GMP plant starts with four outputs in the first 90 days: a walk-down of every asset behind a GMP parameter, a criticality ranking, the calibration status of every BMS and EMS input, and ownership of open HVAC deviations and CAPAs.

InsightRead