Swabs are step four of a cleaning validation. Most arguments over a swab result were settled, or not, in steps one to three.
The sequence below follows section 10 of EU GMP Annex 15 (2015 text).
- Choose the worst case. Where a worst-case product is the model, the Annex says a scientific rationale should be provided for the choice; solubility, cleanability, toxicity and potency are criteria it may rest on.
- Set the limit. Carryover limits should be based on a toxicological evaluation, and the Annex points to the EMA guideline on health-based exposure limits.
- Fix where and how to sample. The protocol should specify the locations and the reason for each. Recovery should be shown to be possible from every product-contact material sampled, with every sampling method used.
- Run it. The time before cleaning and the time before next use should be taken into account to define dirty and clean hold times, and the number of runs should rest on a risk assessment.
- Report, then keep confirming. For manual cleaning the Annex says effectiveness should be confirmed at a justified frequency.
On the limit itself, the EMA Q&A of 2018 says the guideline is not intended to be used to set cleaning limits at the level of the calculated exposure limit. We read that figure as the ceiling, with the working limit below it.