Skip to content

Life Sciences

21 CFR 11.100: what comes before the first electronic signature

  • Life Sciences
  • 1 min read

Before the first record is signed electronically, 21 CFR Part 11 asks the company for a certification signed by hand.

It sits in § 11.100, the general requirements for electronic signatures, and it is the last of three.

Two of the three are about accounts. Each electronic signature shall be unique to one individual and shall not be reused by, or reassigned to, anyone else. A shared "operator" login fails this on day one. So does handing a leaver's account to the replacement. And before the organization establishes, assigns or certifies an individual's electronic signature, it shall verify the identity of the individual.

The certification goes to the agency. Persons using electronic signatures shall certify, prior to or at the time of such use, that the signatures in their system are intended to be the legally binding equivalent of traditional handwritten signatures. It is signed by hand and submitted in electronic or paper form.

All three are duties of the company, and a supplier's "Part 11 ready" data sheet cannot do them for you. They belong in the URS and the site procedure. Whether Part 11 applies at all depends on the market and the predicate rules.

We design automation and SCADA to 21 CFR Part 11 and Annex 11.

Talk to an engineer: contact@humiley.com · +84 28 7777 6668 (Viet Nam office)

Industries

Related

More on this subject

Infrastructure qualification under a GMP SCADA

Servers, network, virtualisation, backup and time sync under a GMP system are qualified as a controlled platform; the application on top is validated for its intended use. With the platform qualified, a routine OS patch needs an impact assessment and a targeted check, not a new OQ of the SCADA.

InsightRead

Part 11 starts with the predicate rule

Part 11 applies to electronic records that a predicate rule such as 21 CFR 211 requires. For a BMS or EMS, name the rule behind each point: GMP-critical monitoring usually belongs in a validated EMS. For EU export, Annex 11 has no predicate-rule test, so the BMS still needs a risk assessment.

InsightRead

Individual logins make records attributable, but user lists drift as people move and leave. A periodic access review exports users and roles from the system, checks them against HR's leavers and movers, justifies every admin or generic account and is signed with its corrections, the record an inspector will ask for.

InsightRead