Calibration before OQ: installed transmitters count too

  • Life Sciences
  • 1 min read

An OQ report can hold a certificate for every test instrument and none for the room pressure transmitters it reads.

Those transmitters are often calibrated at the vendor's factory before shipping. By OQ that can be months earlier, with a wet season and a site installation in between. If one has drifted, the pressure cascade in the report is a figure nobody can defend.

EU GMP Annex 15 includes calibration of instrumentation in the IQ checks. 21 CFR 211.68(a) wants automatic and electronic equipment routinely calibrated or checked, and 211.160(b)(4) bars instruments that fail their calibration specifications.

The test kit has a trap of its own. A particle counter calibrated last year can fall due halfway through the OQ window, and every count taken after the due date becomes a deviation.

Before the first test, one sheet can cover both groups:

  1. Test instruments brought in: particle counter, anemometer or capture hood, micromanometer, data loggers.
  2. Installed sensors the protocol reads: room differential pressure, temperature and RH transmitters, filter gauges.

For each item, record the certificate number, the due date checked against the last day of the test window, and the calibrated range checked against the value to be tested. Put that sheet at the front of the OQ.

Industries

Related

More on this subject

Changing a terminal HEPA while the suite keeps running

Replacing one failed terminal HEPA in a live Grade C suite starts with change control and an equivalent filter. Stop work, bag the old filter, inspect the housing, then fit and leak-scan the new filter in place. Clean and check airflow, pressures, counts and recovery before release, with the scope agreed in the change record.

InsightRead

EU GMP Chapter 3 and 21 CFR 211 on HVAC

21 CFR 211.42 and 211.46 and EU GMP Chapter 3 ask for adequate ventilation, air control and filtration, but give no air change rate or pressure differential, and no filter class beyond HEPA for aseptic areas. Those come from Annex 1, FDA's 2004 aseptic guidance, ISO 14644 and the designer's justification, recorded in the design basis.

InsightRead

Grade B change rooms: one room or two

EU GMP Annex 1 calls separate entry and exit change rooms for Grade B desirable. One room saves area but relies on procedure at the busiest shift change; two rooms give one-way flow at the cost of doors, air and qualification. On a new build, start from two.

InsightRead