The full wet clean comes after batch twelve of the campaign. The cleaning validation was run after batch three.
Campaigns make sense: between batches of the same product a vacuum and a wipe, because a full wet clean after every batch would cost a shift. The trouble is what the equipment looks like at the end. Product bakes onto the warm surfaces of a dryer or coater a little more with each batch. Fines pack into the filter housing and behind seals. Anything damp has had a week to grow. The final clean has to cope with all of that, and it was only ever proven against a lighter load.
The 2015 text of EU GMP Annex 15 has a clause for it, 10.9: where campaign manufacture is carried out, the impact on the ease of cleaning at the end of the campaign should be considered, and the maximum length of a campaign, in time and/or number of batches, should be the basis for cleaning validation exercises.
So the cleaning validation master plan states a maximum campaign, in batches and in days, and a validation run is performed after a campaign of that length. The batch record needs a field for the batch's position in the campaign, so that batch thirteen cannot start unnoticed.