Nearly every instrument list carries a 12-month calibration interval, and few sites can say where the figure came from.
The 2015 text of EU GMP Chapter 3 says measuring, weighing, recording and control equipment should be calibrated and checked "at defined intervals by appropriate methods". 21 CFR 211.68 says equipment shall be routinely calibrated, inspected or checked according to a written program. Neither names a period.
The nearest thing to a figure is WHO's storage and transport guidance: monitoring devices in a store calibrated at least once a year, unless otherwise justified. A storage default, with its exception written in.
The calibration laboratory cannot settle it either. Under ISO/IEC 17025 a certificate shall not carry a recommendation on the calibration interval unless that was agreed with the customer.
That leaves the interval with the site, and it needs a basis. Two inputs do most of the work: what a wrong reading from this instrument would let through, and the as-found results of past calibrations. An instrument found well inside tolerance several times running can go longer; one found near the limit twice should come in sooner.
Legal metrology is a separate obligation: national rules may fix a verification cycle for some instrument types.
Most of the list will stay at 12 months. What changes is one line of rationale per instrument class in the calibration procedure, and the few instruments that move.