The position of the Grade A sampling head decides what a continuous-monitoring result proves.
The limit does not move while the line runs: 3,520 particles ≥ 0.5 µm per m³, at rest and in operation. Continuous monitoring is how you show it held through the batch. Annex 1 (2022) asks for Grade A monitoring at locations that represent the risk to the exposed product. In practice the sampling head belongs close to the point of fill, in air that has just passed the open containers, facing into the airflow. A head mounted behind a machine guard, or downstream where the process has already been cleared, samples air that never came near the vials.
The tubing matters as much. Annex 1 also sets a Grade A monitoring limit of 29 particles ≥ 5.0 µm per m³, and particles that size settle out in long runs and tight bends. A counter in the technical area at the end of several metres of tube can undercount exactly that size. Keep the run short and straight, and ask the supplier for transport-loss data on the installed route.
Test the position rather than assume it. Run the smoke study with the sampling head in place and check on the video that the air reaching it has passed over the open product.
For us the Contamination Control Strategy is the organizing document, and the monitoring points belong in it next to the airflow design that justifies them.