Out of tolerance at calibration: the look-back, then the adjustment

  • Life Sciences
  • 1 min read

Adjust an out-of-tolerance transmitter and close the work order, and nobody asks how long it was wrong or what was decided on its readings.

Say a room pressure transmitter reads 3 Pa high at its annual calibration, against a ±1 Pa tolerance. For up to twelve months, a room displayed at 12 Pa may have been at 9 Pa. The adjustment is step five:

  1. Record the as-found values and tag the instrument. Nothing is adjusted yet.
  2. Find the last calibration that passed. That date opens the window in which the readings are in doubt.
  3. List what relied on the instrument since then: batches made in the room, alarms that stayed silent, a requalification that used its trend.
  4. Raise a deviation and have QA assess the impact. Direction matters: a transmitter reading high hid low pressure, while one reading low mostly caused false alarms.
  5. Adjust or replace, record the as-left values, then review the interval.

EU GMP Chapter 3 says measuring and control equipment should be calibrated and checked at defined intervals, with adequate records. 21 CFR 211.68 says automatic, mechanical and electronic equipment shall be routinely calibrated, inspected or checked according to a written program. Both stop there. For APIs, ICH Q7 adds the look-back: calibration deviations on critical instruments should be investigated for impact on what was made since the last successful calibration.

What does your calibration procedure put between as-found and adjust?

Topics

Industries

Related

More on this subject

Pass boxes: where the grade boundary leaks

A pass box sits across a grade boundary designed for at least 10 Pa. The interlock does not fix a worn door seal, a static box is made safe by the disinfectant contact time rather than by waiting, and a dynamic box needs its supply balanced against both rooms and its HEPA filter leak tested.

InsightRead

Periodic requalification: what is retested and how often

EU GMP Annex 1 (2022) sets maximum requalification intervals: 6 months for Grade A and B cleanrooms and 12 months for Grade C and D. Classification, HEPA leak tests, airflow volume, pressure differentials and air velocity, or recovery where airflow is not unidirectional, are retested. A risk assessment can shorten these intervals, never extend them.

InsightRead

In a Li-ion dry room at −60 °C dew point and below, extra air changes do not help: return air often passes back through the desiccant rotor, so more supply air means a bigger rotor and more regeneration heat. Size airflow from the moisture balance; pressure control, airlocks and envelope sealing carry over from pharma.

InsightRead