Periodic requalification: what is retested and how often

  • Life Sciences
  • 1 min read

A common misreading: "risk-based requalification" means you can stretch the interval if nothing has gone wrong. It works the other way.

EU GMP Annex 1 (2022) sets maximum intervals: 6 months between requalifications for Grade A and B cleanrooms, 12 months for Grade C and D. A risk assessment can justify testing more often, or adding tests, where the process or the monitoring history calls for it. It is not a route to longer gaps in a sterile suite.

What a periodic requalification retests:

  1. Classification by particle count to ISO 14644-1.
  2. Filter integrity: installed HEPA leak test by aerosol scan.
  3. Airflow volume.
  4. Pressure differentials between adjacent rooms and grades.
  5. Air velocity: required for filling zones under unidirectional airflow, and set by the CCS risk assessment in Grades B, C and D. In rooms without unidirectional airflow, a recovery test replaces the velocity test.

The calendar is not the only trigger. A filter replacement, an AHU modification or a layout change goes through change control and may need partial requalification whatever the date is.

Our maintenance team plans requalification from the same asset register and criticality assessment as the rest of the maintenance program, so the test dates don't live in a separate spreadsheet.

Industries

Related

More on this subject

Changing a terminal HEPA while the suite keeps running

Replacing one failed terminal HEPA in a live Grade C suite starts with change control and an equivalent filter. Stop work, bag the old filter, inspect the housing, then fit and leak-scan the new filter in place. Clean and check airflow, pressures, counts and recovery before release, with the scope agreed in the change record.

InsightRead

What ends the life of a HEPA filter

EU GMP Annex 1 (2022) asks for an integrity test of final filters at requalification and puts no age on the filter. A terminal HEPA is replaced when a leak cannot be repaired within site limits, when pressure drop reaches the manufacturer's final value, or after damage or wetting. On glass-fibre media, dust loading alone does not lower efficiency.

InsightRead

Why cleanroom airflow is called unidirectional

At 0.45 m/s across a 1 m clean zone the Reynolds number is about 30,000, far beyond laminar flow. ISO 14644-1 and the 2022 EU GMP Annex 1 use the term unidirectional airflow: steady velocity, roughly parallel streams. It is tested by air speed, uniformity and smoke studies, so write it that way in the URS.

InsightRead