A new pack size on a line that is already qualified

  • Life Sciences
  • 1 min read

Packing a 60-count bottle on a line qualified for the 30-count is a change to the product range.

Same tablet, a taller bottle. Five steps, in order:

  1. Open a change control, with a regulatory assessment. Whether a variation is needed is for regulatory affairs to decide.
  2. Check format parts and settings. A taller bottle brings new guides, a new count setting, another torque setting and a different label position. Compare each with the range the line was qualified for.
  3. Validate where the range moved. When the new settings sit inside the qualified range this step is small, and someone still records that, with the reason.
  4. Release artwork, codes and master data: label, carton, the pack code in the ERP, camera recipes.
  5. Update the packaging instructions, the line clearance checklist and operator training.

Steps 1, 3 and 5 rest on GMP text. Under EU GMP Annex 15 (2015 text), written procedures should describe the actions when a change is proposed to the product range, and quality risk management should be used to evaluate the impact on validation and regulatory status. In the packaging section of the same 2015 text, qualification of primary packing equipment should be carried out at the minimum and maximum operating ranges defined for the critical process parameters. EU GMP Chapter 4 (2011 text) adds that approved Packaging Instructions should exist for each product, pack size and type.

Industries

Related

More on this subject

A technology transfer package needs the reasons with the recipe

A batch record followed to the letter does not tell a receiving site why each parameter sits where it does. The WHO guideline on technology transfer says the sending unit should make available all the necessary information and knowledge. Check for the development history, the justified control strategy, validated analytical procedures with stability data, and hold time and cleaning information.

InsightRead

Engineering Solution & Consultancy

An SOP register answers the change-control question

An SOP register should record, for each SOP, the owner, version, review date and every document it refers to. Humiley's template covers 536 SOPs across 20 departments. The cross-reference column tells a change control which documents a revision touches, and review dates keep superseded versions out of use.

InsightRead

The scope of EU GMP Annex 1 (2022) says some of its principles, including the contamination control strategy, may be used for certain non-sterile products such as liquids, creams and ointments. A semi-solids plant that writes a CCS tying room grades, flows, water and monitoring to product risk has an answer for the inspector.

InsightRead