Reading ISO 29463 HEPA labels against EN 1822

  • News
  • 1 min read

ISO 45H on a HEPA label is 99.995% written as a code: four nines, then a five.

ISO 29463 names filter classes by efficiency at the most penetrating particle size. ISO 35H means at least 99.95% and ISO 45H at least 99.995%, the same integral efficiencies as EN 1822 H13 and H14. That is why the two pairs are treated as equivalents in most specifications.

The class in between is where a swap can slip through. ISO 40H, at 99.99%, sits between H13 and H14 and has no EN 1822 name. Written as penetration, 40H passes 0.01% at MPPS and 45H passes 0.005%, so a 40H filter passes twice as many particles at that size as the H14 it may be offered against. A submittal that reads "ISO 40H, HEPA" is true on both counts and still below your specification.

For the specification, write the EN 1822 class your URS already uses, name the ISO 29463 class you accept as its equivalent, and ask for the individual test report with each filter's serial number. A substitution then has to arrive as a deviation, with a signature on it, instead of as a new label.

Which filter standard do your purchase orders name: EN 1822, ISO 29463, or both?

See how we work

Topics

Related

More on this subject

Receiving HEPA filters: what to check before fitting

A HEPA filter leaves the factory tested, then travels by container, forklift and scaffold. Check the cartons on arrival, match label class, serial number and rated airflow to the schedule and test report, store dry within the stacking mark, open only at a clean work area and inspect media and gasket under strong light before fitting.

InsightRead

Seven HVAC documents, indexed before the inspection

Cleanroom inspectors ask for the same HVAC evidence: qualification report, pressure trends and alarm log, HEPA leak tests, classification and recovery, smoke studies, as-builts and change controls. Index them per room, with document number, revision and location, so the back room answers in minutes.

InsightRead