The inspector never saw your HEPA scan. They read the report, and the report is where installed leak tests quietly fail.
A test can be done well and recorded badly. Before you sign, check seven lines:
- Upstream concentration measured and written down, not assumed. With a photometer, a leak is a reading above 0.01% of it, so without that number the result has no reference.
- Challenge uniformity across the filter face. A poorly mixed challenge makes half the filter look better than it is.
- Probe size and scan rate recorded, and matching the calculation in the protocol.
- Overlapping strokes. Gaps between passes are untested media.
- The frame-to-housing seal scanned, not only the media. The gasket is a leak path a media-only scan never sees.
- Instrument calibration in date on the day of the test, with certificate numbers.
- Filter serial numbers matched to the drawing positions, so a repair or a replacement can be traced to the right filter.
Items 5 and 7 are the ones most often missing. A report that says "all filters passed" without saying which filters, where, is a report you cannot defend.
Periodic requalification is part of our Maintenance Service, and this list is how we read the reports before they go to QA.
Which of the seven does your current contractor leave out?