Forty SOPs signed "read and understood" in one afternoon is a signature log. It says little about what the person can do.
Reading works for some procedures. A document numbering SOP can be learned from the page. Changing a terminal HEPA filter or responding to a differential pressure alarm cannot, because the skill is in the hands and in the order of the steps.
EU GMP Chapter 2 says training should be appropriate to the duties assigned, continuing training should be given, and its practical effectiveness should be periodically assessed. A signature under "read and understood" records that reading happened.
For a practical task the training file needs more. Anything done with tools or in a gown gets a demonstration and supervised practice. The trainee then performs the task while the trainer observes against the SOP and signs the assessment. The file also carries a date for the next assessment.
All of this costs trainer hours, which is the real reason read-and-understood survives. Sorting the SOPs by risk first puts the observed assessments on the tasks where a mistake reaches the product.
How many SOPs on your matrix have an observed assessment?