The HEPA housing joint the filter scan misses

  • Manufacturing & Distribution
  • 1 min read

Around every ceiling HEPA housing there is a second joint, housing flange to ceiling panel, and the installed filter scan never looks at it.

The scan under ISO 14644-3 covers the filter media, its frame and the seal between filter and housing. The flange that holds the housing in the ceiling lies outside it, and a different trade made it on site.

Whether it matters depends on which way the air wants to go. Under a positive room with a return-air void, leakage runs out of the room and costs some pressure. Where the void is at higher pressure than the room, as over a negative containment suite or a pressurised plenum, unfiltered void air comes in beside the filter. The room's particle counts then carry the evidence without pointing at the cause.

Sealing it is simple work done carefully: a continuous gasket under the flange, compressed evenly by the fixings, and a cleanroom-grade sealant bead at the visible edge, finished so it can be wiped.

Checking it takes a particle counter and a few minutes per housing. With the system running, move the probe slowly along the flange and compare with the room background; a smoke pencil shows direction. File both with the filter scan.

HEPA boxes and housings are part of the AeroSano range, built by qualified partners under Humiley engineering control and FAT-tested before they ship.

Related

More on this subject

ULPA U15 to U17: where the extra nines pay

EN 1822 ULPA classes add a nine each above H14: U15 99.9995%, U16 99.99995%, U17 99.999995% at MPPS. Semiconductor ISO Class 1 to 3 spaces need them. A pharma Grade A zone at ISO 5 rarely does, because people and seal leaks drive its counts; H14 with a sound seal usually carries it.

InsightRead

Why cleanroom airflow is called unidirectional

At 0.45 m/s across a 1 m clean zone the Reynolds number is about 30,000, far beyond laminar flow. ISO 14644-1 and the 2022 EU GMP Annex 1 use the term unidirectional airflow: steady velocity, roughly parallel streams. It is tested by air speed, uniformity and smoke studies, so write it that way in the URS.

InsightRead

As-built classification on a new cleanroom suite

ISO 14644-1 allows a room to be classified as-built: installation complete, services running, no equipment, materials or people. On a new suite it shows whether the HVAC, filters and envelope can reach the class alone. It says nothing about machines or people, so the GMP classification is still done at rest and in operation.

InsightRead