The first audit trail challenge at EMS OQ is a plain one: change a single setpoint and read back the entry it created.
21 CFR Part 11 section 11.10(e) asks for secure, computer-generated, time-stamped audit trails, and says a change must not obscure the value recorded before it. An entry reading only "setpoint changed" has lost the old value. EU GMP Annex 11 adds that the reason for a change or deletion of GMP-relevant data should be documented, so the entry needs old value, new value, user, time and reason.
Next, log in as an operator and then as a supervisor and look for any way to pause, edit or purge the trail. If either can, the people whose data it records control the record. Administrator rights belong to someone outside production and QC, and that account's own actions are audited too.
The time stamp is easy to miss. Move the workstation clock forward an hour, make a change, and see which time the entry carries. It should come from a clock users cannot set, usually a server synchronized to a network time source.
Archive readability cannot be shown on the day of OQ. Trails must stay readable for the retention period, including after a software upgrade, so the upgrade plan needs a step that opens an old trail.
We write the URS and validation strategy alongside the design, where these points are cheapest to settle.