18.2 MΩ·cm: the ceiling for pure water at 25 °C

  • News
  • 1 min read

No treatment plant can deliver water above 18.2 MΩ·cm at 25 °C. Water with nothing in it still conducts that much.

A small fraction of water molecules split into hydrogen and hydroxide ions on their own, and those ions carry current. At 25 °C that gives pure water a resistivity of 18.18 MΩ·cm, a conductivity of 0.055 µS/cm. Purification can approach that figure and never pass it, so ultrapure water specifications for semiconductor work read 18.2.

The USP stage 1 conductivity limit for Purified Water at 25 °C is 1.3 µS/cm, about 24 times the conductivity of pure water.

The figure belongs to 25 °C. Pure water reads about 32 MΩ·cm at 15 °C and about 9 at 40 °C, so the meter compensates, and a display above 18.2 points to the compensation or the cell constant. Read it in the line, too: a sample in an open beaker absorbs carbon dioxide and its resistivity falls within minutes.

At this ceiling, resistivity sees ions and nothing else. Organics, particles and bacteria pass under it, so a UPW specification goes on to total organic carbon, particle counts, dissolved oxygen and silica.

Related

More on this subject

AMC limits: name the compound, then the unit

HEPA and ULPA filters do not stop airborne molecular contamination, and an AMC limit written only as "1 ppb" is a different mass for every compound: about 0.70 µg/m³ for ammonia, 1.49 µg/m³ for hydrogen chloride. A fab URS should name the compound, the unit with its reference temperature, and the sampling method.

InsightRead

UHP piping: the pipe is a cleanroom too

High-purity PVDF joined like ordinary plastic pipe leaves fusion beads that trap bacteria, so fab UPW loops use bead-and-crevice-free joints. UHP gas lines add electropolished 316L, orbital welds with coupons, helium leak tests and outlet particle, moisture and oxygen checks. Pharma WFI practice shares much of the discipline.

InsightRead