Count particles before the filter leak test and you may have to throw the result away.
If the scan then finds a leak, the filter is repaired or replaced, and the classification has to be run again on the room as it now stands. Cleanroom OQ tests depend on each other, so their order is part of the protocol.
- Measure airflow volume and velocity first, and finish the balancing. Everything after this assumes the design air quantity.
- Set and record the pressure differences between rooms with all doors closed. Any later rebalancing moves them.
- Scan the installed filters for leaks at the airflow now fixed. Repairs happen here, before anything counts particles.
- Run the airflow visualization. It shows whether the air goes where the design says, and of all the tests it is the likeliest to send you back to step 1.
- Classify the room, then run the recovery test. They come last because the aerosol from the leak test and the smoke study has to clear first.
ISO 14644-3 gives the methods for the supporting tests in steps 1 to 5; the classification itself is to ISO 14644-1. ISO 14644-3 does not set the order: it says its tests are not presented in chronological order, so the sequence belongs in the site's OQ protocol.
Smoke studies are often filmed late, because the camera and QA are booked for one day, and that puts step 4 after step 5. It works only if nobody expects the video to move a diffuser.