Many EM programs still put settle plates on the ISO 14644-1 classification grid, which spreads sampling points evenly by floor area.
Monitoring should go where contamination would reach an open product or a critical surface first. Annex 1 (2022) asks for monitoring locations to come from a documented risk assessment, and the smoke study is one of the best inputs to it.
Five criteria for placing particle counter probes and settle plates:
- Airflow pattern. The smoke study video shows where first air meets the product and where it turns. Probes sit on the path to the open containers.
- Interventions. Plates go where gloves, tools and operators enter Grade A during routine and corrective interventions.
- Proximity without interference. Close enough to represent the critical zone, without disturbing unidirectional flow or the operator's reach.
- Worst case in the background room. Near doors, transfer hatches and returns, where the room recovers last.
- Evidence. Each location on the drawing carries a reference to the smoke study frame, risk assessment line or intervention that put it there.
Airflow visualization is part of the cleanroom qualification Humiley carries out. If a smoke study is still to come, film it with the EM plan in mind, so a probe location can cite the frame showing first air at the filling needles.
Which of your current EM points would be hardest to explain to an inspector?