HEPA leak scans: check the upstream challenge first

  • News
  • 1 min read

Every HEPA leak scan passes or fails against 0.01% of a number that many test reports never print: the upstream concentration.

With a photometer, the aerosol measured upstream of the filter is taken as 100%, and a leak is any steady downstream reading above 0.01% of it. ISO 14644-3 writes the acceptance as that ratio, so a pass is worth only as much as the upstream measurement behind it.

The photometer method works with an upstream challenge of roughly 10 to 100 µg per litre of air. At 20 µg/L the leak threshold is 0.002 µg/L downstream. If the generator runs low, the threshold falls with it: every tenfold drop in the challenge puts the threshold ten times closer to the instrument's zero. A pinhole in the media or a lifted gasket still leaks, but its plume gets hard to tell from background, and the filter is signed off.

We see thin challenges most on large filter banks, where one generator is asked to feed too many filters, and on duct-mounted housings where the upstream sample point sits in air that has not mixed yet.

On a report we want the measured upstream figure for each filter or bank rather than the generator setting, with the point where it was sampled and the photometer's resolution written next to it.

Related

More on this subject

Installing terminal HEPA filters: the sequence

A terminal HEPA filter passes its factory test and is then carried through a building site. Clean the room and run on pre-filters first, unpack at the point of use, check the label against the position schedule, tighten evenly, then leak-scan the installed filter to ISO 14644-3 and record the serial number against its position.

InsightRead

ISO Class 1 vs ISO 5: counting at 0.1 µm

ISO Class 1 allows 10 particles of 0.1 µm and larger per cubic metre; ISO 5 allows 100,000 at that size. At that level the installed leak scan must prove frames and grid, not just the U15 filter, single-digit counts need large sample volumes, and airborne molecular contamination needs its own sampling plan.

InsightRead

EU GMP Annex 1 (2022) sets maximum requalification intervals: 6 months for Grade A and B cleanrooms and 12 months for Grade C and D. Classification, HEPA leak tests, airflow volume, pressure differentials and air velocity, or recovery where airflow is not unidirectional, are retested. A risk assessment can shorten these intervals, never extend them.

InsightRead