Pharmacopoeial methods still need verification

  • Life Sciences
  • 1 min read

A pharmacopoeial HPLC assay was validated before it was published, and it has never seen your placebo.

That is the gap behind "it's a compendial method, we can use it as it is". EU GMP Chapter 6 says testing methods should be validated, and that a laboratory using a method it did not originally validate should verify the appropriateness of the testing method.

WHO's validation guideline goes further. When pharmacopoeial methods are used, evidence should be available to prove that they are suitable for routine use in the laboratory. Methods for content or impurities should also have been demonstrated to be specific for the product under consideration, with no placebo interference. And it describes verification as partial validation: the characteristics are selected according to the procedure and its intended use, and the selection should be justified.

Passing system suitability on the day is a narrower thing. It shows the instrument and column were working for that run.

How much verification is enough depends on the method. A pH test needs far less than a related-substances method on a product with five excipients.

Industries

Related

More on this subject

Receiving-site readiness, one signed document at a time

An assessor reads what stands behind the registration batches. By the time the first one is made, five things should be true at the receiving site: equipment and rooms qualified, analytical procedures transferred, specifications and suppliers approved, the master batch record issued and operators trained. Each needs a signed document behind it.

InsightRead

Recovery test or clean-up period in the URS

The ISO 14644-3 recovery test times a 100:1 fall in particle concentration after an aerosol challenge and measures the air system. The Annex 1 clean-up period, under 20 minutes as a guidance value and set during qualification, describes the room after real operations. Write the recovery test into the URS and set the clean-up period at in-operation qualification.

InsightRead

When an environmental monitoring system is replaced, years of records move with it or become unreadable. List the records and retention periods, specify what moves, verify a trial migration, migrate under change control and keep the old data readable. The 2011 text of Annex 11 says validation should include checks on value and meaning.

InsightRead