Most calibration certificates are filed unread. Check that the laboratory is accredited for this measurement, that tag, serial number and calibrated points match the instrument and its range, read as-found before as-left, compare the uncertainty with the tolerance, and look for traceability and the decision rule behind any pass statement.

InsightRead

Pilot scale for tablets starts at 100,000 units

ICH Q1A(R2) defines a pilot scale batch for solid oral dosage forms as generally at least one-tenth of production scale or 100,000 tablets or capsules, whichever is the larger. Two of three primary stability batches should reach it. At 500 mg a tablet that is 50 kg of blend, which sets the equipment size.

InsightRead

Five daily checks on a negative-pressure isolation room

A pressure monitor that reads the same value for a month may be right, or stuck. The CDC guidelines ask for monitoring preferably daily, with documentation. A short round covers the monitor and its alarm, smoke at the closed door, the door closer and latch, the anteroom doors and the exhaust fan.

InsightRead

Bracketing and matrixing: two different reduced designs

ICH Q1D (2002) describes two reduced stability designs. Bracketing tests only the extremes of a factor such as strength or container size, at all time points. Matrixing tests a selected subset of all the combinations at a given time point. Either should be justified, and either can end in a shorter shelf life.

InsightRead

EU GMP Annex 15 (2015) says a formal release for the next qualification stage should be authorised, and that conditional approval to proceed can be given where there is a documented assessment of no significant impact on the next activity. That assessment is made item by item, against the tests that follow.

InsightRead

OOS and OOT: two results, two different triggers

An out-of-specification result falls outside the specification or acceptance criteria and starts with a laboratory assessment. An out-of-trend result is inside the limits but breaks the pattern of earlier results. EU GMP Chapter 6 says both should be addressed and investigated, which needs data recorded so that trends can be evaluated.

InsightRead

Concurrent validation is the exceptional route

EU GMP Annex 15 (2015) accepts concurrent validation in exceptional circumstances, where there is a strong benefit-risk ratio for the patient. The decision must be justified, documented in the validation master plan and approved by authorised personnel. A project that is running late does not meet that test.

InsightRead

How to plan a smoke study

EU GMP Annex 1 (2022) says airflow pattern studies should be performed at rest and in operation, with video recordings retained. A usable study starts as a plan: locations and interventions from the risk assessment, defined runs, fixed camera and smoke positions, a script with acceptance criteria per shot, and a QA review.

InsightRead

Two mixers of the same volume are not equivalent if one tumbles and the other uses a ribbon. The FDA's SUPAC guidance speaks of equipment of the same design and operating principles, of the same or a different capacity. WHO's guidelines on technology transfer say the comparison should cover working principle, capacity, make and model.

InsightRead

Leave time between SOP approval and effective date

An SOP that becomes effective the moment it is signed puts the next shift in deviation before anyone is trained. The 2011 text of EU GMP Chapter 4 says the effective date should be defined; the gap after approval is where training, the swap of controlled copies and the register update take place.